CONNECT TO FI BER UNDER TEST (FUT)
All fibers and connectors should be cleaned and inspected prior to connection, as described on page 1.
The OCC is typically connected to the FUT via a Monitor Port or splitter/tap to not disrupt service. The
OCC can also be directly connected to the common port or common fiber via a coupler during turn up or
maintenance:
Figure 3: Connecting to Monitor Port
Figure 4: Connecting to Common Cable
1.
Inspect the OCC port on top of the test set.
2.
If connecting to the Common Cable, connect the Common Cable to an optical coupler with the same
connector type and inspect the fiber end face in the coupler.
3.
If connecting to the Monitor Port or Common Port on the Mux, inspect the Mux Port.
4.
Inspect the fiber end face of the Launch Cable.
5.
Connect the Launch Cable to the OCC port.
6.
Inspect the other fiber end face of the Launch Cable.
7.
Connect the Launch Cable to the coupler or Mux port.