Run Test:
1.
Tap the
Real Time
soft key
to start the
test.
2.
After auto-configuration, the OTDR will
perform a connection check to ensure that
the connection is
Good
. If the Connection is
Bad
, disconnect the launch cable, and
reconnect as described on pages 1 and 2,
cleaning every end-face that fails the
inspection test.
3.
The OTDR will perform real-time acquisitions
at the configured wavelength.
4.
Tap the magnifying glass icons to zoom in
, zoom out
, or auto-zoom
the
display. You can also pinch and zoom with
your fingers.
5.
Tap the
Measurements
soft key
to
start Loss, ORL, or Reflectance
measurements.
6.
Tap the
Test
roll down menu at the bottom
left screen to select a measurement type:
Loss
,
ORL
, or
Reflectance
.
7.
Tap and move the
A
and
B
cursors to change
measurement values.
•
Loss
and
ORL
are measured between the
two cursor positions.
•
Reflectance
is measured at the position
of the
B
cursor.
8.
Figure 10
shows
Loss
measurement,
Figure
11
shows
ORL
measurement, and
Figure 12
shows
Reflectance
measurement.
9.
Tap the
STOP
soft key
to stop the Real
Time measurement.
10.
Tap the
Real Time
soft key
to restart
real time measurement.
Figure 8: Expert OTDR
Figure 9: Connection Check & Real-Time Acquisition
Figure 10: Trace View, Loss Result
Figure 11: Trace View, ORL result