Chapter 6
Scope
Configuring the P5000i Scope
User Manual
76100000200 Rev. 001
37
Configuring the P5000i Scope
Scope
connection
1
Plug in your Viavi scope into a USB port from the 6000A V2 Plat-
form.
2
Push the button
S
YSTEM
3
Validate the
Scope
function
.
4
Connect probe with the fiber being inspected.
You may select this option while others are selected (e.g. OTDR).
Configuring the
Scope
1
Press
S
ETUP
key to configure the test.
The following screen displays:
Test
Profile
On the line
Profile
, select the Profile which will be used for the test of
fiber connector:
–
SM_UPC: Pass/Fail criteria for single-mode UPC connectors from
IEC 61300-3-35 standard.
–
SM_APC: Pass/Fail criteria for single-mode APC connectors from
IEC 61300-3-35 standard.
–
SM_PC: Pass/Fail criteria for single-mode PC connectors from IEC
61300-3-35 standard.
–
MM_: Pass/Fail criteria for multimode connectors from IEC 61300-3-
35 standard.
Profiles contain the analysis parameters by which Pass/Fail criteria are
determined.
Fig. 15
P5000i Scope Setup