34
OLTS-85
5 T
IER
1 L
OSS
T
EST
O
PERATION
T
IER
1 L
OSS
T
EST
2.
Tap the
[
More
]
button.
3.
Tap the
[
Length Unit
]
button.
4.
Tap the desired length unit.
To activate or deactivate the connection tone:
√
The instrument is in
Tier 1 Test
mode.
1.
Press the
[
A
]
key.
2.
Tap the
[
More
]
button.
3.
Tap the
[
Enable Tone
]
button to activate or deactivate the
connection tone.
To recall stored measurement data:
For detailed information, see “Memory Management” on
To recall information relating to built-in sources:
√
The instrument is in
Tier 1 Test
mode.
1.
Press the
[
A
]
key.
2.
Tap the
[
More
]
button.
3.
Tap the
[
Info
]
button.
Changing settings for the Tier 1 loss test
Here, the setting can be made to add new, edit, or delete test
setups. Each
Test Setup
can be saved under its own name. The
factory setting is named
Default
.
Note:
Test setups can only be created, deleted, or edited if the
instrument is configured as a
Local
instrument.
The basic setting in Tier 1 loss test is the
Test Type
. The
Test
Type
ca be set to
Loopback
or
Remote
. Depending on your
instrument hardware configuration the device ca be configured
for single-mode or multi-mode measurements. In
Loopback
mode you need only one instrument configured as
Local
. The
measured cable pair needs to be looped with a jumper cable on
the far end while in
Remote
mode the far end needs to be
terminated with an OLTS-85(P) configured as
Remote
.