Viavi OneAdvisor-1000 Manual Download Page 3

VIAVI//Public

OneAdvisor-1000 Portable Network Tester

OneAdvisor-1000 Portable Network Tester

Q U I C K   C A R D

3

C O N F I G U R E   T E S T

Figure 7: Setup, Display

1.

Tap the 

Setup

soft key    

2.

Tap 

Display 

and set 

Distance Unit 

to your 

desired unit of measure.

3.

Tap 

Acquisition

and configure the following 

basic settings for the trace:

Laser: 

Select the wavelength(s) to test.

Select 

All

or select 

1310 nm 

and 

1550 nm 

to enable bend detection. 

Acquisition Mode

:  

Select 

Manual

to manually to enter 

Range and Pulse width settings. 

Select 

Auto

to automatically configure 

Range, Pulse width, and Resolution 
settings.

Select 

SmartAcq

to perform an 

acquisition with a short pulse width, 
followed by an acquisition with a longer 
pulse width.  The first acquisition allows 
events at the beginning of the fiber to 
be detected more precisely. 

Acq. Time

:  Select Realtime, Auto or the 

desired time to sample the fiber.

Launch Cable

: Select and enter 

Length

if you are using a Launch Cable.  
Enter 20 meters or 66 feet for Viavi 
supplied 20-meter launch cables.

4.

Tap 

Alarms

and configure pass/fail criteria:

Set 

Alarm Level 

to 

Fail.

Set 

Threshold

to 

Default

, or select 

alternate alarm thresholds (TIA-568.3, 
User, etc.)

Figure 8: Setup, Acquisition

Figure 9: Setup, Alarms

Summary of Contents for OneAdvisor-1000

Page 1: ...run inspection test Repeat until it passes Figure 2 Inspect Before You Connect Figure 1 Equipment Requirements 4100 series Expert OTDR This quick card explains how to connect to a fiber under test configure EXPERT OTDR test settings using Quick Setup run tests and analyze results on a VIAVI ONA 1000 equipped with a 4100 series OTDR module ONA 1000 equipped with the following Fiber Optics Software ...

Page 2: ...of the Launch Cable 3 Connect the Launch Cable to the OTDR port 4 If the interface to the FUT is a patch cord connect the patch cord to an optical coupler 5 Inspect the FUT connected to the coupler or OPP port and the fiber end face of the Launch Cable 6 Connect the Launch Cable to the coupler or OPP port Figure 5 Connecting to OPP LAUNCH TEST 1 Press the Power button on the top of the OneAdvisor ...

Page 3: ...lect Auto to automatically configure Range Pulse width and Resolution settings Select SmartAcq to perform an acquisition with a short pulse width followed by an acquisition with a longer pulse width The first acquisition allows events at the beginning of the fiber to be detected more precisely Acq Time Select Realtime Auto or the desired time to sample the fiber Launch Cable Select and enter Lengt...

Page 4: ...b at the left screen top to select SmartLink view 2 Set the toggle switch at the right screen top to Event Line 3 View SmartLink results The FUT is displayed as a series of icons representing events front connector launch cable connectors splices bends etc and distances Swipe left or right to view additional icons on spans with more than 4 events The center of the display shows summary results for...

Page 5: ...he upper display Graphical results dB versus distance are shown in the center display A color coded Event table is shown in the lower display Swipe up and down to view all events Figure 13 Trace view Figure 14 Table view 1 Tap the Table tab at the left screen top to select Table view 2 Set the toggle switch at the right screen top to the center off position 3 View table results Distance Loss and R...

Page 6: ...utions com contact 1 844 GO VIAVI 1 844 468 4284 FAST REPORT 1 Tap the File soft key and tap the Project folder 2 Tap the Create Directory soft key and enter a name for your new folder All trace files will be saved to this directory 3 Tap the Exit soft key 4 Tap the Fast Report soft key and enter Cable ID Fiber Number Locations Direction and Save Mode 5 Tap Save 6 Tap to accept the default filenam...

Reviews: