No, select correct input connector. If persists, proceed to next step.
•
Yes, proceed to next step.
3
Is UUT properly configured to receive signal?
•
No, correct UUT configuration. If problem persists, contact TAC.
•
Yes, contact TAC.
No input signal received on the Spectrum Analyzer.
1
Is input cable properly connected to RF Input or RF Duplex Connector?
•
No, properly connect the cable. If problem persists, proceed to next
step.
•
Yes, proceed to next step.
2
Is the RF input connector to which the input cable is connected selected as
the Spectrum Analyzer Input source?
•
No, select correct input connector. If problem persists, proceed to
next step.
•
Yes, proceed to next step.
3
Are RBW and Span settings set to values that provide optimal sweep time?
•
No, adjust settings to obtain optimal sweep time. If problem persists,
proceed to next step.
•
Yes, contact TAC.
Table 8-2 Troubleshooting Guide (Continued) (Sheet 4 of 4)