Chapter 8
Care, Maintenance, and Troubleshooting
Troubleshooting Information
Rev. 001
CX300 ComXpert Operation Manual
22130634
8 - 12
February, 2022
•
No, set Receiver Port to correct input. If problem persists, proceed to
next step.
•
Yes, proceed to next step.
2
Is cable properly connected to selected connector (RF Input or RF
Duplex)?
•
No, properly connect cable. If problem persists, proceed to next step.
•
Yes, proceed to next step.
3
Is the Receiver Frequency to correct frequency?
•
No, set Receiver Frequency to correct frequency. If problem persists,
proceed to next step.
•
Yes, proceed to next step.
4
Is test set Receiver Bandwidth set appropriately for input signal type?
•
No, set Bandwidth correctly. If problem persists, proceed to next
step.
•
Yes, proceed to next step.
5
Is test set Receiver Reference Level set appropriately for the input signal?
•
No, set Receiver Reference Level correctly. If problem persists,
proceed to next step.
•
Yes, proceed to next step.
6
Is Frequency Reference set to Internal?
•
Yes, remove any external source connected to the REF Output
Connector. If Problem persists, proceed to next step.
•
Yes, proceed to next step.
7
Is Frequency Reference set to Internal?
•
No, connect a valid external Reference. If Problem persists, proceed
to next step.
•
Yes, proceed to next step.
8
Is the correct external reference source selected (System > Clock
Source)?
•
No, select correct external reference source. If problem persists,
contact TAC.
•
Yes, contact TAC.
No generator output at RF Output or RF Duplex Connector.
1
Is correct output connector. If problem persists, proceed to next step.
Table 8-2 Troubleshooting Guide (Continued) (Sheet 2 of 4)