Chapter 8
Care, Maintenance, and Troubleshooting
Troubleshooting Information
22130634
CX300 ComXpert Operation Manual
Rev. 001
•
No, select correct output connector. If problem persists, proceed to
next step.
•
Yes, proceed to next step.
2
Is RF Generator output turned ON?
•
No, turn on RE Generator output. If problem persists, proceed to next
step.
•
Yes, proceed to next step
3
Is Generator Frequency set to correct frequency?
•
No, set RF Generator Frequency set to correct frequency?
•
Yes, proceed to next step
4
Is RF Generator level set appropriately for UUT?
•
No, set RF Generator level to correct setting. If problem persists,
contact TAC.
•
Yes, contact TAC.
No input signal received at Audio Input Connector.
1
Is input cable properly connected to Audio Input Connector?
•
No, properly connect the cable. If problem persists, proceed to next
step.
•
Yes, contact TAC.
2
Is the Audio Input Connector to which the input cable is connected selected
as the Audio Input source (Audio Input 1 or Audio Input 2)?
•
Yes, proceed to next step
3
Are Audio in settings appropriate for the characteristics of the incoming
audio signal (e.g. Range, Coupling)?
•
No, adjust settings. If problem persists, contact TAC.
•
Yes, contact TAC.
No signal output at AF Output Connector.
1
Is CX300 AF Generator enabled?
•
No, turn on AF Generator. If problem persists, proceed to next step.
•
Yes, proceed to next step.
2
Is RF input connector to which the input cable is connected selected as the
Spectrum Analyzer Input source?
Table 8-2 Troubleshooting Guide (Continued) (Sheet 3 of 4)