Chapter 6
Performing Tests and Measurements
Channel Power Measurements
Rev. 001
CX300 ComXpert Operation Manual
22130634
6 - 20
February, 2022
6.6.4 Configuring Equipment
The following procedures illustrates the Hardware Setup Diagram.
Connect the CX300 and UUT as shown in
, then proceed to the next section.
The following procedure describes how to configure the Spectrum Analyzer.
In this example, the CX300 Spectrum Analyzer uses the following settings:
1.
Power on the CX300.
2.
Select
Spectrum Analyzer Test
mode.
3.
Select the Full Test Settings button (
) to open the
Full Test Settings Menu
.
4.
Navigate to the
Input/Output
controls and settings.
5. Input Source
: select
RF Duplex
.
6.
Navigate to the
Frequency
controls and settings.
7.
Set the frequency range over which the instrument will sweep.
8.
Navigate to
Trace
controls and settings.
9. Detector Type
: select
RMS
.
10. Average Count
: set to desired value.
11.
Navigate to the
Sweep
controls and settings.
12. Sweep Mode
: select
Continue
.
13. Sweep Type
: select
Auto
.
Figure 6-9 Channel Power Measurements Setup Diagram