Chapter 5
Test and Measurement Functions
Spectrum Analyzer
Rev. 001
CX300 ComXpert Operation Manual
22130634
5 - 68
February, 2022
5.15.1.5 Setting Sweep and Triggers
Sweep settings are used to define the length of time that the instrument takes to sweep
the displayed frequency span (or, in zero span, the time that the instrument takes to sweep
the full screen). To perform measurements using certain signal conditions, use trigger
settings to configure specific trigger events. When using a trigger source, the CX300 starts
to sweep only with the selected trigger conditions are met. The test set can be configured
to user an external or internal trigger source.
Table 5-40 Spectrum Analyzer Band Sweep and Trigger Controls and Settings
Control/Setting
Description
Sweep Mode
This toggle switch selects whether the analyzer performs signal
sweeps continuously, or if the analyzer performs a single
sweep before being switched off.
Single
In single sweep mode, the measurement is stopped after a
single signal sweep.
Continue (Cont)
In continuous mode, the analyzer measures continuously,
repeating the signal sweep as long as the analyzer is ON.
Sweep Type
This setting is not applicable to Zero-Span Analyzer Mode.
Manual
When set to
Manual
, the
Sweep Time
setting is used to
define the Spectrum Analyze sweep time.
Auto
When set to
Auto
, the instrument optimizes the sweep time
based on the characteristics of the incoming signal.
Sweep Time
The
Sweep Time
field is enabled when
Sweep Type
is set to
Manual
.
The
Sweep Time
defines how quickly trace data is acquired
(the length of time an analyzer takes to sweep the defined
frequency span) and to update the display.
This setting must be fast enough to provide quick measurement
results, but slow enough to allow the power values at each
point to be measured. Reducing the sweep time increases the
sweep rate.
Trigger Source
Zero Span Trigger Source
is enabled when
Analyzer Mode
is
set to
Zero-Span
. This menu selects the signal source that the
CX300 uses for trigger events.
Free Run
When
Free Run
is selected triggering occurs immediately
after the a measurement sweep is started.