5.12 Frequency List
The Frequency List tool allows users to define frequency list tables which can easily be
loaded and applied via the CX300 UI. The CX300 supports.csv formatted files. File
contents must be in the order shown below:
Generator Frequency, Receiver Frequency, Generator State, Generator Level
Frequency List pane
is accessed from the Quick Access Toolbar by pressing the
Frequency List button
. When this button is pressed, the
Quick Access Toolbar
updates
to display a pane that contains the frequency list controls and settings.
The
Frequency List pane
displays stored frequency list files. Files can be stored locally
in the test set’s hard drive or accessed from an external storage device. The Frequency
List pane contains the following controls and settings:
Table 5-33 Frequency List Content
List Component
Description
Open File
Opens file browser window to selected frequency list
file. When a file is selected, the RF Generator and/or
Receiver update to the settings defined in the loaded
frequency list file.
Previous
Loads the previous frequency defined in the frequency
list file.
Next
Loads the next frequency defined in the frequency list
file.
Index Button
Displays a frequency from the frequency list file.
Name Button
Displays or selects a frequency from the frequency list
file.
Generator Frequency
Frequency to which RF generator will be set.
UOM: MHz
Receiver Frequency
Frequency to which RF receiver will be set.
UOM: MHz
Set Button
Loads the selected frequency settings.
Generator State
On/Off state of the RF generator
Text Format: On or Off.
Generator Level
Output level to which RF Generator will be set.
UOM: dBm