JD700B Series User’s Guide
771
To set the multiple segment line:
1
Press the
LIMIT
hot key.
2
Press the
Multi Segment Line
soft key for Pass/Fail indication.
3
Toggle the
Limit
soft key between
Upper
and
Lower
to select the one to be displayed.
4
Set the number of segments for the selected upper or lower limit line, up to 50 segments.
Press the
# of Line
soft key.
Enter a value between 1 and 50 by using the numeric keys.
Press the
Enter
soft key.
5
Press the
Autoset
soft key to let the instrument set the limit for each segment and display the line.
6
Optional
. To move the limit line, complete the following steps:
Select
Limit Up/Down
or
Limit Left/Right
.
Turn the rotary knob to move the line as desired. You can also manually enter a value.
7
Optional
. To edit the segment properties, complete the following steps:
Press the
Edit Limit
soft key.
Press the
Move
soft key and then turn the rotary knob to select the segment to edit.
Select the menu option, from the following choices:
To hide the line for the selected segment, toggle the
Line
soft key and select
Off
.
To add a new point, press the
Add Point
soft key.
To delete the selected point, press the
Delete Point
soft key.
To change the position, press the
Frequency
or
Amplitude
soft key, and then turn the
rotary knob to change the value as desired.
8
Optional
. Go to
SAVE/LOAD > Save
, and then select
Limit
to save the limit settings.
See “Using save” on page 35 for more information.
To set the ICNIRP limit:
1
Press the
LIMIT
hot key.
2
Toggle the
ICNIRP Limit
soft key to
On
to show a reference line for the set span frequency
range based on the ICNIRP guidelines or to
Off
to hide the line from the display.
Recording and replaying EMF logging data
Logging EMF spectrum data
You can record EMF spectrum measurement as a logging file in
.grr
file type so that you can play it back in
the spectrogram chart view for data analysis.
Procedure
1
Press the
SAVE/LOAD
hot key.
2
Select the
Save
soft key.
3
Select the
Save Logging Data
soft key.
The on-screen keyboard appears.
4
Type the name of the file in the
File Name
field.
NOTE
In is recommended that you use an external USB memory drive to save logging data in
order to have enough free space.
5
Press the
Save
soft key to start logging.
The “
Recording
” appears on the screen.
6
Press the
ESC
hard key to stop logging data and complete saving.