Using RFoCPRI Analyzer
328
Selecting port
CellAdvisor 5G supports two antenna ports for signal transmission, and each port can generate up to four different waveforms
selectively and simultaneously. User can select the port to use for the measurement in any measurement modes of RFoCPRI mode.
1
Tap the
Setup (
)
icon on the side bar of each measurement mode and choose
Rx Settings
or
Port Configuration
.
2
Select
Port 1
or
Port 2
in
Select Port
line.
Configuring test parameters
Configuration of test parameters described in this section is used in the RFoCPRI mode including spectrum, spectrogram and
spectrum replayer measurements.
Setting frequency
You can set the frequency with either frequency or channel number. If a frequency to be set matches to the frequency
corresponding to the selected channel standard, the instrument calculates its channel number and updates the screen with it
automatically.
To set the frequency with center frequency:
1
Tap
Frequency
under
Menu
(
).
2
Tap
Center Frequency
and enter a value by using the on-screen keyboard.
3
Select a unit from
GHz, MHz, kHz
, or
Hz
and tap
Apply
.
To set the frequency with channel number:
1
Tap
Channel
under
Menu
(
).
2
To select the standard channel:
a
Tap
Channel Standard
, then the standard channel window appears.
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......