5G TM Signal Analyzer
314
Setting measure setup
Before starting the measurement, you need to set Spectrum measurements displayed on the quick access and display tab. See
"
Configuring Spectrum Measurements
" for more details. The measurement settings can be saved and recalled as a file by selecting
File Type as Setup and load the file onto the instrument using the
Save/Load
icons on the side bar.
1
Tap the
Setup
(
) icon on the side bar.
2
Tap
Radio Setup
then the Radio Setup pop-up window appears, if you choose
FR1
, select from the following:
a
SCS 15 kHz: 5, 10, 15, 20, 25, 30, 40, 50 MHz
b
SCS 30 kHz: 5, 10, 15, 20, 25, 30, 40, 50, 60, 70, 80, 90, 100 MHz
c
SCS 60 kHz: 10, 15, 20, 25, 30, 40, 50, 60, 70, 80, 90, 100 MHz
d
Tap
Duplex Type
between
FDD
and
TDD
as required.
-
FDD
: both uplink and downlink can transmit at the same time at different spectrum frequencies
-
TDD
: both uplink and downlink use the same spectrum frequencies but at different times
3
Tap
Radio Setup
then the Radio Setup pop-up window appears, if you choose
FR2
, select from the following:
-
SCS 60, 120 kHz: 50, 100 MHz
-
You can only select the
Duplex Type
as
TDD
.
4
Based on the above setting, tap
Test Model FR1
and select the options from the following choices: FR1-TM1.1, FR1-TM1.2,
FR1-TM3.1, FR1-TM3.1a, FR1-TM3.2, FR1-TM3.
5
Based on the above setting, tap
Test Model FR2
and select the options from the following choices: FR2-TM1.1, FR2-TM2, FR2-
TM3.1.
6
Tap to switch
Phase Correction
to
On
or
Off
.
Phase Correction is a way of optimizing EVM performance. When you switch it On, it will decrease inter-symbol interference
and compensate distortion caused by a filter applying to each OFDM symbol.
7
Tap
Symbol Phase Comp
and select from the following choices:
Auto
,
Manual
, or
Off
.
Symbol Phase Compensation is used to compensate for phase differences between symbols caused by upconversion. Users
do not always use the instrument based on RF frequency. In this case, you can set it to Manual or Off.
▪
Auto
: Setting radio frequency to center frequency.
▪
Manual
: Setting radio frequency as required.
▪
Off
: Setting radio frequency to 0.
NOTE:
When calculating Symbol Phase Compensation, you can use the following formula. It is a formula that complies
with 3GPP standard and used to calculate conversion of 5G NR signal to RF signal.
8
Tap to switch
Select Mode
to
Frame
or
Slot
.
If Slot is selected, tap the
Slot No
. from 0 to 19 using the on-screen keyboard. You can also set it on the screen at the bottom.
Setting limit
1
Tap the
Menu
>
Limit
on the side bar.
2
Tap the test items and set the limits depending on your selected measurement mode:
To set the limit
for
Select
Set
Frequency error
Frequency error
Test limits On/Off, High Limit/Low Limit
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......