CAA06M User's Guide
21
O-S-L calibration using electronic EZ-Cal kit
The e-Calibration is an Open-Short-Load calibration using the electronic EZ-Cal kit, which allows you to
perform the O-S-L calibration easy and fast with only one connection of the kit to the instrument.
Procedure
1. Connect an extension cable, if necessary, to the
Reflection / RF Out
port of the instrument.
2. Connect the
EZ-Cal
to the instrument's
Reflection / RF Out
port or to the extension cable if used.
3. Connect the USB cable to the side of the
EZ-Cal
and then to the
USB Host
port to power the e-
calibration kit.
4. Tap
Cal
icon in the right side bar. The on-screen instruction appears to guide you through the e-
calibration.
5. Tap the
Continue
button. The calibration progress bar appears.