Microstream CO2 Messages
Messages – Measurement Mode
Message
Description
INITIALIZATION
Initializing the EtCO2 Module when power on. It doesn’t measure
during this time.
FILTERLINE NOT
CONNECTED
The filter line is not connected to the EtCO2 accessory.
AUTO ZERO IN PROGRESS
Zeroing the EtCO2 Module by itself. The EtCO2 waveform and
value are not displayed during this time.
It will start auto-zeroing after starting, and at 5, 10, 15, 30, 60
minute intervals and after that, every 60 minutes. During the
warming-up (minimum 20 minutes), it starts auto zeroing < 15
seconds, if it is necessary. Auto zeroing starts at over 8
◦
C or over
5mmHg of pressure change from last zeroing.
CO2 VALUE OVER RANGE
The CO
2
value is out of range.
Range: 0- 99mmHg / 0-13.2% / 0-13.2kPa
MESUREMENT MODE
Wait for EtCO2 measurement with connecting to the accessory.