Apolix

Electronic and automatic polymerizer

Rev.  10/2019

Instruction for use

a.gree srl

Via Retrone 14, Altavilla Vicentina 36077 (VI) Italy

+39 0444 960 991   [email protected] 

1/7

vertysystem.com

1. DESCRIPTION

This instrument is used to polymerize acrylic material. It was engineered for work in the dental laboratory.

1.  Front panel

2.  ON/OFF switch

3.  Manometer

4.  Lid

5.  Handle for closing/opening

6.  Nozzle for water

7.  Nozzle for air

8.  Socket

9.  Power cord

10.  Fusibles

11.  Pressure adapter

12.  Latch for closing/opening

13.  Filter

14.  Temperature button

15.  Timer button

16.  Water release button

17.  RUN/STOP button

18.  Selection arrows

19.  Display

2. INSTALLATION AND FIRST USE

OPENING 

•  Carefully remove the machine from the wrapping

•  Make sure the electric tension is 230 V

•  Set the machine so that the plug is accessible

SET UP 

Connect the main air to the terminal labelled INPUT AIR on the back panel. The input hole has a 6 ø. Pressure must not 

exceed 6 bars.

Connect the 8 ø hose to the OUT output and unwind it to a water drain or a water collecting container. This hole is on the 

left side of the side lid. 

IMPORTANT

Do not soak this hose into any liquid as to not damage the unit. The safety valve is on the back panel. 

Put the instrument in such a position as to not damage anything if the valve is activated. Do not direct it towards 

combustibles.

 

3. MODE OF USE AND CHARACTERISTICS

•  APOLIX has a total of ten programs, four of which are pre-set for VERTYSYSTEM® acrylic

•  materials

•  A work cycle consists of the necessary time and temperature settings

•  Use the selection arrows (e) to set timing and temperatures

•  After using the selection arrows, you will see the current setting flashing on the display (f)

•  When the setting is flashing, you can change its value by using the selection arrows (e)

•  To save the setting, you must once again push the corresponding selection button

•  You will then exit the parameter setup mode

Summary of Contents for Apolix

Page 1: ...the main air to the terminal labelled INPUT AIR on the back panel The input hole has a 6 Pressure must not exceed 6 bars Connect the 8 hose to the OUT output and unwind it to a water drain or a water...

Page 2: ...10 C it s necessary to hold the RUN STOP button 2 for a few seconds to start a new cycle The new cycle will start after an acoustic signal Pressure will stay active until you manually release it Do no...

Page 3: ...onsumption 1220 W 3 Tension category II 4 Actual chamber volume 1 5 l 5 Inside dimensions 140 h 145 mm 6 Maximum pressure 6 bars 7 Working temperature limits 40 C 120 C 8 Temperature and functioning a...

Page 4: ...e at the same time for 3 seconds f Push the RUN STOP button to confirm 8 Regulate temperature t a Hold the temperature button a and set the desired temperature using the arrows e b Push the temperatur...

Page 5: ...d set the second desired temperature using the arrows e b Push the temperature button a again to confirm 10 Regulate duration time du a Hold the timer button b and set the desired time using the arrow...

Page 6: ...arrows e at the same time for 3 seconds The display f will say AIR 3 The pressure will be the same as the last used program 4 Release the air pushing the RUN STOP button temperature duration 2 tempera...

Page 7: ...programme for heat curing full denture resins charged with composite micro pearls For a very high quality polymerization set the pressure to 2 Bar and fill the chamber with at least 3 4 of water Ideal...

Reviews: