Apolix
Electronic and automatic polymerizer
Rev. 10/2019
Instruction for use
a.gree srl
Via Retrone 14, Altavilla Vicentina 36077 (VI) Italy
+39 0444 960 991 [email protected]
1/7
vertysystem.com
1. DESCRIPTION
This instrument is used to polymerize acrylic material. It was engineered for work in the dental laboratory.
1. Front panel
2. ON/OFF switch
3. Manometer
4. Lid
5. Handle for closing/opening
6. Nozzle for water
7. Nozzle for air
8. Socket
9. Power cord
10. Fusibles
11. Pressure adapter
12. Latch for closing/opening
13. Filter
14. Temperature button
15. Timer button
16. Water release button
17. RUN/STOP button
18. Selection arrows
19. Display
2. INSTALLATION AND FIRST USE
OPENING
• Carefully remove the machine from the wrapping
• Make sure the electric tension is 230 V
• Set the machine so that the plug is accessible
SET UP
Connect the main air to the terminal labelled INPUT AIR on the back panel. The input hole has a 6 ø. Pressure must not
exceed 6 bars.
Connect the 8 ø hose to the OUT output and unwind it to a water drain or a water collecting container. This hole is on the
left side of the side lid.
IMPORTANT
Do not soak this hose into any liquid as to not damage the unit. The safety valve is on the back panel.
Put the instrument in such a position as to not damage anything if the valve is activated. Do not direct it towards
combustibles.
3. MODE OF USE AND CHARACTERISTICS
• APOLIX has a total of ten programs, four of which are pre-set for VERTYSYSTEM® acrylic
• materials
• A work cycle consists of the necessary time and temperature settings
• Use the selection arrows (e) to set timing and temperatures
• After using the selection arrows, you will see the current setting flashing on the display (f)
• When the setting is flashing, you can change its value by using the selection arrows (e)
• To save the setting, you must once again push the corresponding selection button
• You will then exit the parameter setup mode