SIB71256 

256 Channel MAPMT Interface Board 

Hamamatsu H13700 Series 

Summary of Contents for SIB71256

Page 1: ...User Guide Vertilon Corporation 66 Tadmuck Road Westford MA 01886 Tel 978 692 7070 Fax 978 692 7010 www vertilon com SIB71256 256 Channel MAPMT Interface Board Hamamatsu H13700 Series...

Page 2: ......

Page 3: ...itation loss of use loss or alteration of data delays lost profits or savings arising from the use of this document or the product which it accompanies Vertilon reserves the right to change this produ...

Page 4: ...SIB17256 Sensor Interface Board for Hamamatsu H13700 MAPMT 4 Vertilon Corporation 66 Tadmuck Road Westford MA 01886 Tel 978 692 7070 www vertilon com...

Page 5: ...1886 Tel 978 692 7070 www vertilon com Table of Contents Product Overview 8 Included 8 Typical Setup 9 Specifications 10 Photomultiplier Tube Anode Circuit 11 Anger Logic 12 Last Dynode Output 13 Gain...

Page 6: ...igure 1 Functional Block Diagram 9 Figure 2 Typical Setup 9 Figure 3 Resistive Anger Logic Circuit 11 Figure 4 Position Signal Plot 12 Figure 5 Anode Preamplifier Response 12 Figure 6 Dynode Preamplif...

Page 7: ...rd or damage to the product do not apply a voltage to any input outside of its specified range Electrostatic Discharge Sensitive Electrostatic discharges may result in damage to the SIB71256 For this...

Page 8: ...tor Event position and energy information can be determined by connecting the four anger signal outputs to customer supplied electronics For timing applications utilizing the last dynode output of the...

Page 9: ...data acquisition system Digitized output data from the PhotoniQ is sent through a USB 2 0 connection to a PC for display logging or real time processing The amplified last dynode signal from the SIB7...

Page 10: ...0 ohms Bandwidth 160 MHz Slew Rate 450 V us Output Baseline Level 12 55 50 110 mV mV No load Isink 5mA Maximum Output Pulse 4 V Into 50 ohms Output Impedance 50 LAST DYNODE PREAMPLIFIER Input Resistan...

Page 11: ...ow Each anger signal is amplified by high speed inverting preamplifiers and output on the SMB connectors on the four corners of the printed circuit board The gain of the preamplifiers can be configure...

Page 12: ...241 V16 V256 V1 V16 V241 V256 An actual plot of the position signal using the above formulas is created by successively injecting a negative DC current into each anode and collecting the charge data u...

Page 13: ...determining time of arrival or coincidence detection The figure below shows the last dynode preamplifier response in the high gain setting to a voltage signal injected into the DY10 location on the S...

Page 14: ...Figure 8 Top and Bottom Views 1 Anger SMB Output V16 2 Gain Adjust G16 3 Gain Adjust G1 4 Anger SMB Output V1 5 Last Dynode SMB Output DY 6 Gain Adjust DY 7 Anger SMB Output V256 8 Gain Adjust G256 9...

Page 15: ...er Guide 15 Vertilon Corporation 66 Tadmuck Road Westford MA 01886 Tel 978 692 7070 www vertilon com Mechanical Information Figure 9 SIB71256 Printed Circuit Board Dimensions Figure 10 SIB71256 3D Vie...

Page 16: ...including without limitation loss of use loss or alteration of data delays lost profits or savings arising from the use of this document or the product which it accompanies Vertilon reserves the right...