71
4.
Setting
If you need to set the functions, power on the device, openthe battery cover, press the "SET” button to enter
in function setting mode, continuous operation for the next option, wait for 2 seconds it will enter into the
submenu automatically
The function list as follows:
NAME
FUNCTION
DESCRIPTION
Celsius/ Fahrenheit unit option
Select the temperature unit
Body/Object
temperature selection
Ear/Forehead mode or Ear/Object mode
selection. “body” means body tempearature
mode,“FACE”means object temperature
mode.
Time setting
When you set time,the room temperature
and time will display alternately
Offset setting
Set the temperature offset according to one’s
need as there maybe difference when skin
color difference. The range is ±3°C(for
forehead mode only). If your difference is
negative, press the button until a minus is
displayed in front of the value.
Orange backlight temperature range
setting.
The system default temperature range is
37.5°C -37.9°C, the user defined
range is ±0.3°C
Red backlight temperature range
setting.
The system default temperature range is
38.0°C -42.9°C, the user defined range is
±0.3°C, it will display priority when
temperature range overlap with the range of
orange backlight
Warning ! To save the settings above, you must press the setting button until you see "SAVE" on the
display.
5.
Temperature measurement
When put on the probe cover and the LCD displays “
“, it switch into forehead measuring mode, when
take off the probe cover, it switch into ear measuring mode and the LCD display”
”
1)
Forehead temperature
Power on, point the device to the forehead, Press down the “Power/Scan” button, Keep the probe cover
contacting with the skin, scan it across the forehead from one side to the other side for 2-3 seconds,
Release the “Power/Scan “button, then the device makes a sound “beep” the measurement is completed.
2)
Ear temperature
Power on, take off the probe cover and put the probe in the ear where you feel comfortable, Press the
“Power/Scan” button, then the device makes a sound “beep”, the measurement is completed.
3)
Ambient temperature
Clock
Low battery
Memory of last body temperature
Summary of Contents for INFRACOLOR
Page 1: ...THERMOMETRE MULTISCAN AURICULAIRE FRONTAL INFACOLOR MODE D EMPLOI 70340 0040...
Page 3: ...1 THERMOMETRE MULTISCAN AURICULAIRE FRONTAL INFRACOLOR MODE D EMPLOI 70340 0040...
Page 13: ...11 MULTISCAN OHR STIRNTHERMOMETER INFRACOLOR GEBRAUCHSANWEISUNG 70340 0040...
Page 25: ...23 MULTISCAN THERMOMETER OOR VOORHOOFD INFRACOLOR GEBRUIKSAANWIJZING 70340 0040...
Page 36: ...34 TERM METRO MULTISCAN AURICULAR Y FRONTAL INFRACOLOR INSTRUCCIONES DE UTILIZACI N 70340 0040...
Page 47: ...45 TERM METRO MULTISCAN AURICULAR E FRONTAL INFRACOLOR INSTRU ES DE UTILIZA O 70340 0040...
Page 58: ...56 TERMOMETRO MULTISCAN AURICOLARE E FRONTALE INFRACOLOR MANUALE D USO 70340 0040...
Page 69: ...67 Infrared Thermometer USER MANUAL 70340 0040...