Vernier BlueView Transilluminator How To Use Download Page 1

 

BlueView Transilluminator 

 

(Order Code BLUE-VIEW) 

 

 

and  

Blue Digital Bioimaging System

 

 

(Order Code BL-DBS) 

The Blue View Transilluminator is an 

illuminating base for observing fluorescently 

stained nucleic acid and protein gels. Using a 

filtered bright blue LED light source, this 

transilluminator excites stained gel bands 

resulting in light emissions in the visible light 

region of the spectrum.  
Use the BlueView Transilluminator: 

 

To visualize fluorescent bands from gel electrophoresis results. 

 

To safely observe gel digest results. 

 

As part of the Blue Digital Bioimaging System to photodocument and analyze 

gel electrophoresis results. 

 

To analyze transformation results using Green Fluorescent Protein (GFP).  

What is Included with the BlueView Transilluminator? 

 

BlueView Transilluminator 

 

12 V AC Power Adapter 

How to Use the BlueView Transilluminator  

Connect the BlueView Transilluminator to AC power using the cable provided. 

Note:

 Do not use any other power supply with the transilluminator. Lift the lid and 

place the gel or tray holding the gel on the blue platform. Close the lid. Turn on the 

unit by turning the knob and adjust the brightness for optimal viewing.  

How the BlueView Transilluminator Works 

The BlueView Transilluminator uses bright blue 

LEDs and a blue filter to produce light with a peak 

wavelength of 470 nm as shown in Figure 1. This 

wavelength excites nucleic acids and proteins 

stained with certain dyes, causing them to 

fluoresce. The orange lid filters out the blue light 

to reveal only the fluorescently stained bands.  
A rotary dial control switch adjusts light intensity 

to optimize visualization. A magnetic switch turns 
off the LEDs when the lid is open. 

 

Figure 1: Emission spectrum of 
 the BlueView Transilluminator  

 2

Which Stains Can Be Used with the BlueView Transilluminator? 

Fluorescent stains and dyes that work well with the BlueView Transilluminator 

include, but are not limited to, the following.

 
For nucleic acid gel electrophoresis:  

 

SYBR

®

 Safe

 

 DNA Gel Stain  

 

SYBR

®

 Green I DNA Gel Stain 

 

SYBR

®

 Gold DNA Gel Stain 

 

GelStar

®

 Nucleic Acid Gel Stain 

 

GelGreen

 Nucleic Acid Gel 

Stain 

 

Ethidium bromide 

For protein gel electrophoresis: 

 

SYPRO

®

 Orange Protein Gel 

Stain 

 

SYPRO

®

 Ruby Protein Gel 

Stain 

 

Coomassie Fluor

 Orange 

 

BlueView Transilluminator Specifications 

AC power supply to external power brick: Input 0.8 A @ 100 to 240 V AC 
DC power input to transilluminator: 2.0 A @ 12 V 
Dimensions (excluding power supply): 17 cm 

 23 cm 

 10.5 cm 

Care and Cleaning 

For best results, keep the viewing surface as dry as possible. Should the lid or 
viewing surface require cleaning, disconnect from power, use soapy water, and wipe 
dry with a soft cloth to avoid scratching. Do not use solvents, glass cleaners, 
cleansers, or any solution containing ammonia on these surfaces. If liquid wicks 
under the blue viewing surface, disconnect power, unscrew and remove the four 
Phillip’s screws, separate the white and blue plates, wipe dry with a soft cloth, and 
reassemble before connecting power. If the lid fogs when viewing a gel, lift and 
gently wipe with a soft cloth.  

Patent Information 

The BlueView Transilluminator includes Clare Chemical Research, Inc.'s 

Dark Reader

®

 transilluminator technology and is covered by one or 

more of Clare Chemical Research, Inc.'s U.S. Patent nos. 6,198,107, 

6,512,236, and 6,914,250. 

Warranty 

Vernier warrants the BlueView Transilluminator to be free from defects in materials 

and workmanship for a period of five years from the date of shipment to the 

customer. This warranty does not cover damage to the product caused by abuse or 

improper use. 

Note:

 This product is to be used for educational purposes only. It is not appropriate 

for industrial, medical, research, or commercial applications.

 

Summary of Contents for BlueView Transilluminator

Page 1: ...ns Can Be Used with the BlueView Transilluminator Fluorescent stains and dyes that work well with the BlueView Transilluminator include but are not limited to the following For nucleic acid gel electrophoresis SYBR Safe DNA Gel Stain SYBR Green I DNA Gel Stain SYBR Gold DNA Gel Stain GelStar Nucleic Acid Gel Stain GelGreen Nucleic Acid Gel Stain Ethidium bromide For protein gel electrophoresis SYP...

Page 2: ...w the documentation and analysis sequence in the following section Obtaining a Digital Photo There are two ways to obtain a digital photo for Gel Analysis in Logger Pro In one method a photo is inserted into Logger Pro from an existing file on your computer In the second method Logger Pro actively captures a photo of a gel using a digital camera such as the ProScope HR Figure 2 Blue Digital Bioima...

Page 3: ...calculations will be correct 6 Set Standard Ladder In this step you will identify the bands of the standard ladder and input their base pair values Logger Pro will automatically create a standard curve on the graph Click the Set Standard Ladder button Click on the band closest to the well of the standard ladder lane Type the corresponding number of base pairs into the field provided in the resulti...

Page 4: ...erials and workmanship for a period of five years from the date of shipment to the customer This warranty does not cover damage to the product caused by abuse or improper use 8 Vernier Software Technology 13979 S W Millikan Way Beaverton OR 97005 2886 Toll Free 888 837 6437 503 277 2299 FAX 503 277 2440 info vernier com www vernier com Rev 11 5 2014 Logger Pro is our registered trademark in the Un...

Reviews: