Vernacare Compact Plus Technical Manual-
Revision 2-Issued by N.Shields 14.2.2020
Page | 13
4.
Installation / Commissioning
4.1.
Required Install Clearences and Heights
Note: Installation of this equipment should only be carried out by a suitable qualified person.
A minimum of 50mm clearance is required at the back of
the machine. This allows for access to the water inlet and
electrical connection at the rear of the machine.
A minimum of 150mm clearance is required either side of
the Compact. This allows tool access during adding/
removing of the front panel. Servicing and maintenance
procedures may also be carried out without having to
move the machine.
150mm
150mm
50mm
350mm
SIDE & REAR MACHINE CLEARENCE
LID CLEARENCE & MACHINE HEIGHT
A minimum of 350mm clearance is required above the
Compact lid. Items fitted to the wall directly above the
machine may restrict the lid from opening and operating
effectively.
The foot sensor is pre-set to operate at the height of a
floor mounted Compact Plus (range highlighted in green).
If a plinth is installed the sensor will have to be
reprogrammed to adjust the range by a trained engineer.
Summary of Contents for COMPACT+