Vernacare Compact Plus Technical Manual-
Revision 2-Issued by N.Shields 14.2.2020
Page | 40
FAULT
EXPLANATION
OPERATOR CORRECTIVE
ACTIONS
MAINTENANCE CORRECTIVE
ACTIONS
‘POWER FAIL’
Error Message
Displayed
The electricity
supply was
interrupted
Press the start button to
resume the cycle. If the
problem persists call
maintenance
Check electrical supply
Check 24 V DC power supply unit
‘HOPPER FULL’
Error Message
Displayed
There is liquid in
the hopper from a
previous cycle,
possibly due to
something
blocking the drain
outlet
Open lid and check for
liquid in the hopper. If the
hopper does not drain call
maintenance.
Check sensor operates correctly,
recalibrate if necessary
‘LID CANNOT
OPEN’
Error
Message
Displayed
Something is
obstructing the lid
Check nothing is on top of
the lid, press start button to
clear the error. If the
problem persists call
maintenance
Check lid open and lid closed sensors for
correct operation.
Check magnetic clutch for slipping
Check lid motor is working
‘LID CANNOT
CLOSE’
Error
Message
Displayed
Something is
obstructing the lid
Check nothing is
obstructing the lid, press
start button to clear the
error. If the problem persists
call maintenance
Check lid open and lid closed sensors for
correct operation.
Check magnetic clutch for slipping
Check lid motor is working
‘DRAIN VALVE’
Error Message
Displayed
The drain valve
has not closed
properly
Call maintenance
Check pressure sensor for correct
operation.
Check airline connections for leaks
Check Solenoid valve for correct
operation.
Check diaphragm valve integrity