94
VTK-VS-BA-015e-A | User Manual MDS 3200+ Series
|
Cleaning
−
Clean the outside of the tappet centering piece first with a lintless rag. Then use a fluidic
brush for the bore.
−
Finally you clean all O-rings with a lintless rag.
Step 7:
Dry all the components in the air or with compressed air.
In case you still have parts which are not completely clean after the fine purification, repeat
steps 5 and 6 for those components, several times if necessary. If that is still not helping,
contact our Technical Support.
8.4.5
Assembling of the Fluid Box
Step 8:
Re-assemble the valve and its components in the following order.
−
Place the tool MDT 301 upright and push the tappet sealing with the wider side down
onto the pin. Situate the fluid box with the bore over the tappet sealing and push the flu-
id box straight down. A light „click“-noise will tell you, if the tappet sealing sits tight in
the fluid box. You can remove the fluid box from the tool.
−
Press the tappet centering into bore of the fluid box. Make sure it sits plane on the tap-
pet sealing. The picture shows you how deep inside the bore the tappet centering piece
has to sit.
(If you use a tappet sealing LX instead of the pictured tappet sealing, you do
not need a tappet centering piece. Refer to the manual “Mounting Instruction – Tappet
Sealing LX”, which is included on the CD, with details how to mount it correctly.)