![VERDER CARBOLITE GERO Installation, Operation And Maintenance Instructions Download Page 15](http://html.mh-extra.com/html/verder/carbolite-gero/carbolite-gero_installation-operation-and-maintenance-instructions_1003717015.webp)
l
Remove the back panel.
l
Tighten the three hexhead
screws entering through the key-
hole slots to secure the two flow-
meter modules together, taking
care not to damage any internal
tubes or fittings.
l
Replace the back panel.
5.3
Gas Connections
5.3.1
Gas Inlet
l
The gas inlet comprises a 6mm outside
diameter fitting, into which an appro-
priately sized hose can be pushed.
l
Fully insert the hose into the fitting to cre-
ate a secure connection.
l
The other end of the hose should be con-
nected to an external gas supply (cus-
tomer supplied).
Note:
To disconnect any pipes from the gas inlet and push fittings, push and hold
down the blue ring whilst withdrawing the hose.
15