2.0 Installation
Depending on the application of the product, it may be appropriate to position it under
an extraction hood. Ensure the extraction hood is switched on during use.
Ensure that the product is placed in such a way that it can be quickly switched off or
disconnected from the electrical supply.
Under no circumstances should any objects be placed on top of the product.
Always ensure that any vents on the top of the product are clear of any
obstruction. Always ensure all cooling vents and cooling fans (if fitted) are
clear of any obstruction.
2.3
Chimney
The chimney is a length of tubing. If it is supplied unfitted, then fit it through the hole in
the top of the case.
If the product is to be used to heat substances that emit fumes, a fume extraction duct
of approximately 75 mm - 150 mm inlet diameter may be placed directly above the
chimney outlet.
Do not make a sealed connection to the product chimney as this causes excessive
airflow through the chamber and results in poor temperature uniformity.