thermocouple and temperature indicator should be made from time to time to
determine whether full calibration is required. Carbolite Gero can supply these items.
Depending on the controller fitted, the controller instructions may contain calibration
instructions.
6.4
After-Sales Service
Carbolite Gero Service has a team of Service Engineers who can offer repair, calibration
and preventive maintenance of furnace and oven products both at the Carbolite Gero
factory and at customers’ premises throughout the world. A telephone call or email
often enables a fault to be diagnosed and the necessary parts to be despatched.
In all correspondence please quote the serial number and model type given on the
rating label of the product. The serial number and model type are also given on the
back of this manual when supplied with the product.
Carbolite Gero Service and Carbolite Gero contact information can be found on the back
page of this manual.
6.5
Recommended Spare Parts and Spare Parts Kit
Carbolite Gero can supply individual spare parts or a kit of the items most likely to be
required. Ordering a kit in advance can save time in the event of a breakdown.
Each kit consists of one thermocouple, one solid state relay, one door insulation piece
and a set of elements.
When ordering spare parts, please quote the model details as requested above.
6.6
Power Adjustment
The control system incorporates electronic power limiting, but for the model listed in
this manual the power limit is set to 100%. The power limit parameter OP.Hi may be
accessible to the operator, but should not generally be altered.
In some cases the supply voltage may be outside the range 220-240 V or the 3-phase
equivalent, the power limit parameter may be set to a value other than 100%. Do not
increase the value to 100%, see section 10.0 for details of power limit settings.
17
6.0 Maintenance
Summary of Contents for CARBOLITE GERO HRF 7/324
Page 26: ......
Page 27: ...Notes Service Record Engineer Name Date Record of Work ...