![VERDER CARBOLITE GERO ABF 8/28 Installation, Operation And Maintenance Instructions Download Page 39](http://html.mh-extra.com/html/verder/carbolite-gero-abf-8-28/carbolite-gero-abf-8-28_installation-operation-and-maintenance-instructions_1003626039.webp)
materials that emit smoke or fumes, regularly heat it up to maximum temperature for
one hour with the chamber empty to burn away the soot.
The product has been designed for ashing items that containing no more than 40 g of
carbon.
Baskets are provided with a loading handle. These can be used to load and unload the
items being ashed. Make sure a safe surface is available, which must not be flammable,
available to place the basket when unloading.
7.3
Operator Safety
This product incorporates a safety switch which interrupts the heating
element circuit when the furnace is opened. This prevents the operator
touching a live heating element and also prevents the product from
heating up if the furnace is left open. The operation of this switch should
be checked periodically.
Depending on use, the surfaces in the working chamber and the
chamber load may still be very hot after the appliance is switched off.
Touching these surfaces may cause burns. Use suitable personal
protective equipment or wait until the appliance cools down to ambient
temperature.
Before removing a hot object from the product, make sure there is a safe place to put it
down. If necessary use tongs, face masks and heat resistant gloves. Heat resistant
clothing and face protection can guard against the effects of radiated heat when the
furnace is open.
Do not open the product door during an ashing procedure as combustion
will be taking place within the chamber.
7.4
Temperature Control
Main controller Eurotherm 3216 P1 maximum temperature 800 °C. See controller
instructions.
Afterburner controller Eurotherm 2132 maximum temperature 950 °C. See controller
instructions.
Element over-temperature controller 2132. This controller is mounted inside the
furnace and has a set temperature of 900 °C. This controller automatically switches off
the elements when the element chamber gets to hot and then switches back on. The
element over-temperature controller is fitted as additional protection for the heating
elements. This over-temperature is self resetting and should not need any user
adjustment. It is accessed by removing the back panel – see photo 3.5 below:
39
7.0 Operation
Summary of Contents for CARBOLITE GERO ABF 8/28
Page 11: ...Fig 14 Schematic for Mains Electrical Connections 11 3 0 Installation ...
Page 21: ...4 6 Navigation Diagram 21 4 0 2416 Controller ...
Page 33: ...5 7 Navigation Diagram 33 5 0 Afterburner 2132 Controller ...
Page 40: ...7 0 Operation Photo 3 5 40 ...
Page 53: ...Notes Service Record Engineer Name Date Record of Work ...