![VERDER CARBOLITE GERO ABF 8/28 Installation, Operation And Maintenance Instructions Download Page 10](http://html.mh-extra.com/html/verder/carbolite-gero-abf-8-28/carbolite-gero-abf-8-28_installation-operation-and-maintenance-instructions_1003626010.webp)
3.0 Installation
Fig 13 - Ducting Schematic
3.5
Electrical Connections
Connection by a qualified electrician is recommended.
For electrical connection details of the ABF 8/28 refer to drawing 00348-1-5003
Look at the rating label before connection to check the electrical supply voltage and
frequency for which the product is configured. If there is a difference between the
actual supply and the rating label, please call Carbolite Gero.
Either wire directly to an isolator or fit with a line plug. An isolating switch should operate
on all live conductors (three phase); it should be within reach of the operator. A line plug
should be easily removable and should be within reach of the operator.
The electrical supply MUST incorporate an earth (ground).
Failure to wire to the appropriate diagram may damage the ABF.
10