2.4
Fitting the Chimney and Heating Elements
Please refer to the Element Installation and Replacement in section
6.0. Wear eye protection when handling the heating elements.
The silicon carbide elements are EXCEPTIONALLY FRAGILE and are
packed separately, together with other items shown in the list.
Separately Packed Items
BLF
Elements
6
Braids*
Off 3
* or combined clips and braids may be supplied
BLF /3 transit clamps
Upper Clamp:
1. Remove lower clamp first
2. Remove two M8 nuts and washers, supporting
plate (3) from underneath
3. Remove plate
Lower Clamp:
1. Remove bottom M8 nut
2. Remove two M6 bolts
3. Remove plate
4. Unscrew rod, using spanner on (5) if neces-
sary
Transit Clamps – dependent on model: the example shown here is for the 3 litre
models; remove any transit clamps as shown. Fit the ceramic chimney to the hole that
exits the top of the chamber. Most models are supplied with an insulation piece to fit
around the chimney; fit this insulation.
Remove the complete top access panel using the hidden screw in holes in the top of the
panel; loosen the screws two turns and remove the panel.
Fit the heating elements vertically into the chamber. Handle the elements with great
care. The installation of these elements is described in section 0.1.
9