background image

Calibration option

Calibration button

CALIBRATION QUICK REFERENCE

Sign in 

®

ScanPoint

®

 Docking Station

Top (Gender) button

Water line

0900-0621-06-60

1

Prepare Calibration Phantom

Place on flat, non-reflective surface;
fill with clean, room temperature water
to indicator mark; make sure water 
contains minimal amount of bubbles

ScanPoint

®

 Docking Station

Login name

Password

2

 

Place Instrument in ScanPoint

®

 

 

Docking Station

3

 

Log On to ScanPoint

®

 

First enter login name and  

 

password, then click Sign in 

 

button  

5

 

Click the Calibration Button

6

 

Wait While ScanPoint

®

 Prepares Instrument

7

 

When Prompted, Scan Calibration Phantom

 

Place Instrument in lid.  Ensure that black tip of scanhead  

 

is submerged in water.  Then press the Top (Gender) button. 

 

8

 

Wait While Instrument Calibrates

9

 

Return Instrument to ScanPoint

®

 Docking Station

 

10

 

Select ”Print” Message to Print Certificate   

 

 

of Calibration

Calibration

Calibration

Four arrows appear on the LCD screen when 
analysis of the calibration data is complete.   

ScanPoint

®

, Verathon

®

, and Verathon Medical

®

 are either registered trademarks or trademarks of Verathon Inc. in the

United States and/or other countries. Copyright © 2006 by Verathon Inc. All rights reserved.

www.verathon.com   https://my.scanpoint.com

For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical

®

 representative.

4

 

Select “Calibration” on ScanPoint

®

 Menu