Calibration option
Calibration button
CALIBRATION QUICK REFERENCE
Sign in
®
ScanPoint
®
Docking Station
Top (Gender) button
Water line
0900-0621-06-60
1
Prepare Calibration Phantom
Place on flat, non-reflective surface;
fill with clean, room temperature water
to indicator mark; make sure water
contains minimal amount of bubbles
ScanPoint
®
Docking Station
Login name
Password
2
Place Instrument in ScanPoint
®
Docking Station
3
Log On to ScanPoint
®
First enter login name and
password, then click Sign in
button
5
Click the Calibration Button
6
Wait While ScanPoint
®
Prepares Instrument
7
When Prompted, Scan Calibration Phantom
Place Instrument in lid. Ensure that black tip of scanhead
is submerged in water. Then press the Top (Gender) button.
8
Wait While Instrument Calibrates
9
Return Instrument to ScanPoint
®
Docking Station
10
Select ”Print” Message to Print Certificate
of Calibration
Calibration
Calibration
Four arrows appear on the LCD screen when
analysis of the calibration data is complete.
ScanPoint
®
, Verathon
®
, and Verathon Medical
®
are either registered trademarks or trademarks of Verathon Inc. in the
United States and/or other countries. Copyright © 2006 by Verathon Inc. All rights reserved.
www.verathon.com https://my.scanpoint.com
For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical
®
representative.
4
Select “Calibration” on ScanPoint
®
Menu