27
Operations & Maintenance Manual: Maintenance & Troubleshooting
7. In Scan Point
®
, select the device, verify that the serial number matches the device you are calibrating, and
then click the calibration tank icon.
8. In the Calibrate Instrument window, click the
Calibrate
button. Scan Point prepares the device for
calibration.
9. When the text
Make Phantom measurement, return to cradle
is highlighted and the status says
waiting
, remove the device from the Scan Point docking station and place it into the recess in the
calibration tank lid. Ensure that the tip of the scanhead is submerged in the water.