22
CLEANING & DISINFECTING
Failure to follow these instructions may cause device damage not covered by the warranty:
• Do not immerse the device in cleaning or disinfectant solution or other liquids.
• Do not subject any part of the device to steam, ethylene oxide, radiation, or similar methods
of sterilization or autoclaving.
• Do not use metal or abrasive brushes. These may scratch the device, causing permanent damage.
• Do not use CIDEXPLUS
®
to disinfect the device. CIDEXPLUS will damage the plastic enclosure.
CAUTION
This system may only be cleaned and disinfected by using the approved processes provided in
this manual. Cleaning and disinfection methods listed are recommended by Verathon
®
based on
compatibility with component materials.
WARNING
Availability of cleaning, disinfection, and sterilization products varies by country, and Verathon is
unable to test products in every market. For more information, please contact Verathon Customer
Care or your local representative. For contact information, visit
WARNING
Cleaning is critical to ensuring the component is ready for disinfection. Failure to properly clean the
device could result in a contaminated device after completing the disinfection procedure.
WARNING
Ensure that you follow the manufacturer’s instructions for handling and disposing of the
cleaning and disinfection solutions provided in this manual.
WARNING
When preparing and using one of the approved cleaning, disinfection, or sterilization solutions,
follow the instructions of the solution manufacturer. Pay close attention to the proper dilution
and immersion times.
WARNING
Summary of Contents for BLADDERSCAN BVI 6400