71
Operations & Maintenance Manual: Product Specifications
BATTERY CHARGER/WIRELESS HUB SPECIFICATIONS
The battery charger/wireless hub is powered from a standard wall outlet (adaptable to international power
standards). The battery charger/wireless hub can charge two batteries simultaneously.
Table 25. Battery Charger/Wireless Hub Specifications
SPECIFICATION
DESCRIPTION
Operating Conditions
Use
Indoor
Ambient temperature range
5–40ºC (41–104ºF)
Atmospheric pressure range
700 hPa–106 hPa
Relative humidity
30–75% non‑condensing
Water resistance
Rated at IPX 0 (ordinary equipment without protection against ingress
of water)
Computer connection
USB 2.0
Charger
Powered by a desktop DC power supply.
Input voltage
100–240 V AC RMS
Input frequency
50–60 Hz
Input current
1 A max
Input connection
2 wire IEC 60320 C7
Output
9 V at 1 A
Insulation
Class II with double insulation
Fuses
250 V AC, 2 A, quick acting
Testing
CSA 60950‑1‑03/UL 60950‑1
Storage Conditions
Storage
Indoor
Ambient temperature range
20–60ºC (4–140ºF)
Atmospheric pressure range
500 hPa–1060 hPa
Relative humidity
20–95% non‑condensing