Care and Maintenance
NeoGraft 2.0® User Manual
7-13
7.4. Cleaning the Incision Handpiece
7.4.1.
Disconnecting the Incision
Handpiece Components
To disconnect the incision handpiece components:
1.
Remove the handpiece from the system.
2.
Unscrew the handpiece body from the handpiece base.
3.
Remove the blade holder from the handpiece base.
4.
Carefully remove the blade from the blade holder.
5.
Discard the blade in a hazardous sharps disposal container.
7.4.2.
Cleaning the Incision Handpiece
Body and Blade Holder
To clean the incision handpiece body and blade holder:
1.
Soak the components of in an ultrasonic bath, filled with
ANIOS CLEAN EXCEL D 0.5% enzymatic detergent solution
(25 ml for
5 lit of water) at room temperature (15°C - 25°C) for a
minimum of 10 minutes.
2.
Remove the components from the ultrasonic bath and clean
each component with an appropriate brush (e.g. toothbrush or
interdental brush) for a minimum of 20 seconds. Make sure to
brush inside and outside of the components.
3.
Use an interdental brush (Ø 8-7 mm) to clean the internal parts
of the handpiece body for a minimum of 20 seconds.