Maintenance
5.1
Battery Replacement
The largest aspect of VR4-UWM hardware maintenance is replacing the battery pack. To do this, open the
case, remove the battery, insert the new battery, and close the case. These steps are explained in the
following sections.
5.1.1
Opening VR4-UWM case
The VR4-UWM receiver case should
only be opened when the battery must be
replaced.
To conserve battery power, put the receiver in Suspend mode (see
section 3.3.1) when it is being stored.
Inspect the VR4-UWM case for any sign of leakage or pressurization (see section 8.2.2). This
could mean drying the seam completely and watching for drops of water around the seam, or
spreading soapy water around the seam and watching for bubbles indicating air escaping.
STEP 1
Clean the VR4-UWM thoroughly to remove dirt, mud, and marine growth. Rinse it with fresh water to
remove any salt residue.
Only open the
VR4-UWM case when
replacing the battery.
WARNING
It is vitally important that the electronics inside the receiver case
do not come
in contact with any water
or the receiver will be damaged. For this reason,
VEMCO recommends the VR4-UWM be opened in a controlled environment
and
not
in the field or on the deck of a boat.
STEP 2
Completely dry the receiver before attempting to open it. If available,
compressed air may be used to blow water from crevices. Make sure your hands
and tools are clean and dry before proceeding. Any moisture or contaminants
introduced inside the case can cause leaks, corrosion, short circuits, and
malfunctions. Make sure there is
absolutely no water
on or near the VR4-UWM
case.
This also means the electronics cannot be exposed to a
sudden change in
temperature and humidity
that will cause condensation to develop on the
electronics and destroy the receiver.
Continued…