Velp Scientifica VTF F208B0063 Instruction Manual Download Page 6

1.  Introduction 

 

The  Thermoregulator  is  ideal  for  many  applications  and  meets  the  most  demanding 
requirements in terms of precision, reliability and flexibility of use thanks to the application of 
Fuzzy  Logic  technology.  The  Fuzzy  Logic  electronic  automatically  adapts  thermoregulation 
to  the  varying  factors  such  as  power,  load  and  thermal  dispersion  specific  to  each 
application, optimizing both overheating and oscillations around the temperature set-point. 
The  accuracy  and  precision  of  thermoregulation  at  each  end  the  scale  and  whatever  the 
volume being processed, is a fundamental characteristic of the Vertex. 
With the VTF EVO version, it is possible set different cycles of temperatures and record the 
real values, using the wireless technology in combination with A00000244 TEMPSoft™ with 
W

ireless DataBox™. 

WARNING

:  when  the  Thermoregulator  is  connected  to  another  instrument  than  the  VELP 

Scientifica AREX, AREX DIGITAL PRO, AREC.X or PW10, please check the compatibility of 
the DIN 5 pole connector as illustrated in wiring diagram in this manual. 

NOTE

: the most precise results are obtained measuring aqueous solutions temperature. 

 

The microprocessor offers various other functions:  
•  a working time of up to 24 hours and 59 minutes can be set with automatic switch off; 
•  the maximum sample temperatures reached during the test can be recorded. 
The structure is made of a non-scratch technopolymer resistant to chemical agents and   

 

 

        

Fig. 1

 

offers a high level of IP54. 
The  instrument  has  an  in-built  safety  and  control  circuit  which  is  constantly 
active  and  shuts-down  the  thermoregulator  immediately  in  the  following 
situations: the temperature probe is not connected; the temperature probe is 
faulty  (cut-off  or  short-circuited),  the  temperature  is  out-of-range  and/or  the 
probe is not inserted inside  the sample. 

2.  Assembly and installation 

 

Check the integrity of the unit after unpacking. The box includes: 

 

Thermoregulator, with power cable for connection to VELP Scientifica  
AREX, AREX DIGITAL PRO, AREC.X and PW10  

 

Pt100 temperature probe  

 

Instruction manual 

2.1 Connecting the Pt100 probe 

Connect the probe to the Thermoregulator as shown on 

Fig. 2

NOTE: 

the mains cable must remain far away from the hot plate. 

2.2 Installing the VTF 

The  VTF  has  an  innovative  integrated  system  to  simplify 
installation on the support rod and facilitate the positioning of the 
Pt100 probe in the most commonly used containers.  
The two clamps allow height-regulation whilst a sliding  
clamp allows horizontal regulation (

Fig. 3

). 

 

 

 

 

      

Fig. 2 

 

 

        Fig. 3

 

2.3 Using the VTF with magnetic stirrers 

The  power  cable  is  suitable  for  connection  to  the  most  common  heating  magnetic  stirrers  with  a  dedicated  socket  for  the 
remote  control  of  the  heating  plate  such  as  the  VELP  Scientifica  models  AREX,

 

AREX  DIGITAL  PRO  and  AREC.X.  The 

socket  must  supply  a  tension  of  between  9  and  15V  DC.  The  integrated  installation  system  allows  the  instrument  to  be 
installed on support rods with a diameter of between 10 and 13 mm.  

2.4 Using the VTF EVO in wireless mode. 

For  more  information  about  the  wireless  function  of  VTF  EVO,  please  check  the  operation  manual  of  A00000244 
TEMPSoft™ with Wireless DataBox™. 

2.5 Other applications 

When using the thermoregulator for other applications bear in mind that the transistor output can take a maximum current of 
50mA. Where power loads of up to 2200W are required the use of an external power relay is necessary (code A00000001, 
VELP  PW10).  For  the  thermoregulation  of  liquids  that  are  not  compatible  with  the  construction  material  of  the  probe 
(stainless  steel  AISI  316),  the  use  of  a  glass-coated  probe  is  recommended  (code  A00000003).  When  it  is  necessary  to 
thermoregulate a liquid that is not in close proximity to the VTF, a 1m probe extension cable is available (code A00000002). 

EN 

1  Display 

 

 

 

2  Set 

3  Minus 

 

 

      4  Plus 

5  Measurement unit   6  Timer display on 
7  Temp. probe   

 

Green “ON” led  

9  Clamps 

 

 

 

10  Plug 

11  Sliding clamp 

Summary of Contents for VTF F208B0063

Page 1: ...solo per applicazioni di laboratorio La societ produttrice declina ogni responsabilit sull impiego non conforme alle istruzioni degli strumenti Cet instrument ne peut tre utilis que pour des applicati...

Page 2: ...ng der Produktqualit t das Recht auf nderung der Ger teeigenschaften vor Safety Regulations Norme di Sicurezza Consignes de Securit Advertencias de Seguridad Sicherheitshinweise The plug disconnects t...

Page 3: ...3...

Page 4: ...I FUNZIONAMENTO 11 3 1 SELEZIONE DELLA TEMPERATURA DI LAVORO 11 3 2 IMPOSTAZIONI TIMER 11 3 3 ALLINEAMENTO DELLA TEMPERATURA 12 3 4 UNIT DI MISURA 12 3 5 SICUREZZE E SEGNALI DI ALLARME 12 4 AVVIO 12 5...

Page 5: ...ACTER STICAS T CNICAS 21 8 ACCESORIOS REFACCIONES 21 1 EINF HRUNG 22 2 MONTAGE UND INSTALLATION 22 2 1 ANSCHLUSS DES PT100 F HLER 22 2 2 INSTALLIEREN DES VTF 22 2 3 KOMBINIEREN DIE VTF MIT MAGNETR HRE...

Page 6: ...integrity of the unit after unpacking The box includes Thermoregulator with power cable for connection to VELP Scientifica AREX AREX DIGITAL PRO AREC X and PW10 Pt100 temperature probe Instruction ma...

Page 7: ...rom the memory every time the instrument is turned off 3 2 Setting the timer From the main window press the key twice to display the working time Working time From 00 00 to 24 59 h min Default value 0...

Page 8: ...ime required To optimize thermoregulation and improve temperature homogeneity stir the sample gently but continuously WARNING When working with set points near the temperature of evaporation check tha...

Page 9: ...e settings intervals C F 10 300 14 572 1 1 VTF EVO Temperature settings intervals C F 10 400 14 752 1 1 Resolution precision C F 0 2 1 0 5 0 9 Working time settings h min From 0 00 to 24 59 Continuous...

Page 10: ...rmoregolatore VTF con cavo di collegamento ad AREX AREX DIGITAL PRO AREC X e PW10 Sonda di temperatura Pt100 Manuale di istruzioni 2 1 Montaggio sonda di temperatura Pt100 Collegare la sonda completa...

Page 11: ...ualizzer la temperatura letta in quel momento dalla sonda immersa nel liquido e termoregoler il liquido alla temperatura impostata NOTA dalla finestra principale tenendo premuto il tasto possibile vis...

Page 12: ...dalla sonda Il ripristino del normale funzionamento avviene spegnendo e riaccendendo lo strumento dopo aver rimosso la causa d allarme Per disattivare AL 2 e AL 3 accendere lo strumento tenendo premut...

Page 13: ...l alimentazione con un panno inumidito con detergenti non infiammabili e non aggressivi 7 Dati tecnici Alimentazione 12 V 9 15 Potenza W 1 2 Dimensioni mm BxHxP 75x145x120 Peso Kg 0 250 Display LCD a...

Page 14: ...de r glage avant arri re 10 Alimentateur 2 Montage et installation Lors de la r ception et apr s avoir enlev l emballage contr ler que l instrument est int gre La fourniture comprend Thermor gulateur...

Page 15: ...des heures aux minutes appuyez sur une fois ou attendez 4 secondes Fig 7 Fig 7 Avec l affichage clignote utiliser et pour s lectionner le temps de travail Si aucune touche n est actionn e pendant 5 s...

Page 16: ...temp rature d vaporation v rifier que la sonde reste immerg dans l chantillon d au moins 10mm travers le cycle de travail L chantillon est soumis l vaporation et pour cette raison il ne peut pas atte...

Page 17: ...10 300 14 572 1 1 VTF EVO R glages temp rature intervalles C F 10 400 14 572 1 1 R solution pr cision C F 0 2 1 0 5 0 9 Temps de travail h min De 0 00 24 59 Mode continu temps de lecture restant Poss...

Page 18: ...dor VTF con cable de conexi n a AREX AREX DIGITAL PRO AREC X y PW10 Sonda Pt100 Manual de instrucciones 2 1 Montaje de la sonda Pt100 Conectar la sonda completa con tapa protectora como se mostra en l...

Page 19: ...predeterminado 0 00 Fig 6 Fig 6 Con y durante la etapa de visualizaci n intermitente es posible programar las horas del tiempo de funcionamiento Para pasar de horas a minutos pulse una vez o espere 4...

Page 20: ...po de funcionamiento Para un mejor control y uniformidad de la temperatura se recomienda una continua y suave agitaci n PRECAUCI N Si se trabaja con un set point pr ximo a la temperatura de evaporaci...

Page 21: ...C F 10 300 14 572 1 1 VTF EVO Temperaturas selec intervalo C F 10 400 14 572 1 1 Resoluci n precisi n C F 0 2 1 0 5 0 9 Tiempo de funcionamiento seleccionable h min De 0 00 a 24 59 Operaci n continua...

Page 22: ...die Temperatur ist au erhalb des zul ssigen Bereichs und oder die Sonde nicht innerhalb eingef gt die Probe 2 Montage und Installation Bitte berpr fen Sie nach dem Auspacken den einwandfreien Zustand...

Page 23: ...g Im Hauptfenster dr cken Sie zweimal um die Arbeitszeit zu anzuzeigen Arbeitszeit von 00 00 bis 24 59 h min Standardwert 0 00 Abb 6 Abb 6 Wenn der Anzeige blinkt verwenden Sie und um die Arbeitszeit...

Page 24: ...Sie die Magnetr hrer Temperatur auf Maximum W hlen Sie die Temperatur und die Arbeitszeit R hren Sie die Probe sanft aber kontinuierlich WARNUNG Beim Arbeiten mit der Temperatur nahe der Verdampfung...

Page 25: ...ierbare Temperatur Intervalle C F 10 300 14 572 1 1 VTF EVO Programmierbare Temp Intervalle C F 10 400 14 572 1 1 Regelung Genauigkeit C F 0 2 1 0 5 0 9 Programmierbare Heizdauer h min Von 0 00 bis 24...

Page 26: ...osotros casa fabricante VELP Scientifica declaramos bajo nuestra responsabilidad que el producto es conforme con las siguientes normas Der Hersteller VELP Scientifica erkl rt unter eigener Verantwortu...

Page 27: ...27...

Page 28: ...zers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e Mi...