1.  Introduction 

 
Thermo reaction can be carried out quickly and safely by placing the tubes into the thermoreactor heating block. 
Set the working temperature from room to 160 °C and the working time up to 199 minutes, or continuous. 
ECO 16 offers the possibility to thermoreact 14 tubes (Ø 16 mm) and 2 (Ø 22 mm) simultaneously. 

2.  Assembly and installation 

 
Upon  receipt  and  after  having  removed  the  packaging,  please  check  the  integrity  of  the  instrument.  The  box  includes:

 

ECO 16 

 

Power supply cord 

 

Instruction manual 

2.1 Electrical connection 

Before connecting the instrument to the power supply, make sure that the values on the rating plate correspond to those of 
the power supply. Connect the unit to the power supply using the transformer supplied. 
Ensure that the socket and the relative cut-off device conform to current safety norms and easy to reach. 

2.2 Start-up 

Use the on-

off switch on the front of the unit to turn the unit on (position “−”). When switched on the unit all the leds blink for 

few seconds. When the leds are off the unit is ready to be programmed at different work temperatures and times. 

3.  Operating controls 

 
Shift (

A

): it moves the messenger under all the parameters to be set during programming. 

Increase (

B

): it shifts the value of the parameter corresponding to the messenger position. 

Enter (

C

): it stores the values shown by the display. 

Start (

D

): it sets going the work cycle. 

Main switch (

E

): it turns on and off the instrument. 

 
 
 
 
 

A B C D  E 

3.1 Programming 

Turn on the main switch on control panel (E). 
The unit performs in a few second a "Self-Test", after which: VELP SCIENTIFICA    20 °C    *READY* 
Pressing the Shift push-button the display shows two values contained in the memory (temperature and time): 
Set  

 Temp.  Time 

 

100 °C  30' 

Where: 
Temp. (100 °C) is the temperature set. Range from room temperature to 160 °C. 
Time (30') is the time of staying at the temperature set. Range from 0 to 199 minutes. 
A continuous operation can be set by a 200, 300... value of time. A symbol === is displayed in this case. 
The parameters set are stored by pressing Enter (C). When the unit is turned off the set values are retained in the memory. 

3.2 Work cycle 

It is set going by pressing the Start (D) and the display shows: Working   20 °C 

60' 

Where: 
Working 

Shows the operation of the work cycle. 

20 °C   

Is the present temperature of heating block. 

60' 

 

Is the time of staying at the set temperature. 

When the set value of temperature is reached the value of time displayed starts to decrease with a resolution of one second. 
At the end of work cycle a beeper warns and the display shows: *END* 
 
When a new work cycle should start with a temperature of the heating block higher than the set value, the display shows: 
VELP SCIENTIFICA   160 °C   NOT READY 
Where the 160 °C refers to the temperature of the heating block and NOT READY refers to the instrument operation.  
If the Start push-button is pressed the cycle does not start and the display shows: NOT ACTIVE   Temp. too high 
Should a work cycle be interrupted the push buttons (C) and (B) must be pressed contemporaneously. 

3.3 Heating block temperature verification 

The temperature is controlled electronically which ensures stability without temperature oscillations around the set point. 
The  Pt  100  probe  provides  high  precision  and  does  not  require  calibration  because  the  internal  software  performs  a  self-
calibration every time the instrument is turned on. If, in spite of everything, a control is necessary, it is possible to measure 

EN 

Summary of Contents for ECO 16

Page 1: ...reil comme d chet solide urbain conform ment la Directive 2002 96 CE No tirar el aparato en los desechos urbanos como exige la Directiva 2002 96 CE Dieses Ger t unterliegt der Richtlinie 2002 96 EG un...

Page 2: ...pos el ctricos de medici n control y su uso en laboratorio Sicherheitsbestimmungen f r elektrische Mess Steuer Regel und Laborger te IEC EN 61010 1 Electrical equipment for laboratory use UL 61010 1 G...

Page 3: ...VIO 7 3 CONTROLLI DI FUNZIONAMENTO 7 3 1 PROGRAMMAZIONE 7 3 2 CICLO DI LAVORO 7 3 3 VERIFICA DELLA TEMPERATURA DEL BLOCCO 7 3 4 ARRESTI 8 3 5 SICUREZZE ELETTRICHE 8 3 6 OPERAZIONI A FINE LAVORO 8 4 MA...

Page 4: ...NES 12 1 EINF HRUNG 13 2 MONTAGE UND INSTALLATION 13 2 1 ANSCHLUSS AN DAS STROMNETZ 13 2 2 INBETRIEBNAHME 13 3 BEDIENUNGSELEMENTE 13 3 1 SCHEDULING 13 3 2 ARBEITSZYKLUS 13 3 3 BERPR FUNG DER TEMPERATU...

Page 5: ...h button the display shows two values contained in the memory temperature and time Set Temp Time 100 C 30 Where Temp 100 C is the temperature set Range from room temperature to 160 C Time 30 is the ti...

Page 6: ...d e g palletized It is the responsibility of the user to properly decontaminate the unit in case of hazardous substances remaining on the surface or interior of the device If in doubt about the compat...

Page 7: ...splay mostra i valori contenuti nella memoria caratterizzati da temperatura e tempo Set Temp Time 100 C 30 Dove Temp 100 C la temperatura di lavoro programmata Range da ambiente a 160 C Time 30 il tem...

Page 8: ...e procedere alla decontaminazione dell unit nel caso in cui sostanze pericolose rimangano sulla superficie o all interno del dispositivo In caso di dubbi sulla compatibilit di un prodotto per la puliz...

Page 9: ...me 100 C 30 Lorsque Temp 100 C est la temp rature de travail programm e Plage de temp rature ambiante 160 C Time 30 est l heure programm e de la permanence de la temp rature Plage de 0 199 minutes Il...

Page 10: ...L instrument doit tre transport dans son emballage d origine et les indications pr sentes sur l emballage d origine doivent tre suivies par exemple palettis r Il est de la responsabilit de l utilisate...

Page 11: ...memoria Set Temp Time 100 C 30 Donde Temp 100 C es la temperatura de trabajo programada Rango de temperatura ambiente a 160 C Time 30 es el tiempo programado de la permanencia a la temperatura Rango...

Page 12: ...por ejemplo paletizado Es responsabilidad del usuario descontaminar la unidad en el caso de que haya restos de sustancias peligrosas tanto en la superficie como en el interior del equipo En caso de d...

Page 13: ...Set Temp Time 100 C 30 wo Temp 100 C ist die programmierte Temperatur von Raumtemperatur bis 160 C Time 30 ist die programmierte Zeit von 0 bis 199 Minuten bei der gew hlten Temperatur Sie k nnen eine...

Page 14: ...twortung des Benutzers das Ger t ordnungsgem zu dekontaminieren falls gef hrliche Substanzen auf der Oberfl che oder im Inneren des Ger ts verbleiben Wenn Sie Zweifel an der Vertr glichkeit eines Rein...

Page 15: ...o conforme alle seguenti norme Nous VELP Scientifica d clarons sous notre responsabilit que le produit est conforme aux normes suivantes Nosotros casa fabricante VELP Scientifica declaramos bajo nuest...

Page 16: ...yzers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e M...

Reviews: