7
3 Product description
VEGATOR 142 •
46839-EN-190626
3 Product description
3.1 Configuration
The scope of delivery encompasses:
•
Controller VEGATOR 142
•
Documentation
– This operating instructions manual
–
Ex-specific "
Safety instructions
" (with Ex version)
–
Safety Manual "Functional safety (SIL) acc. to IEC 61508" (with
SIL version)
–
If necessary, further certificates
The type label contains the most important data for identification and
use of the instrument:
•
Instrument type
•
Information about approvals
•
Technical data
•
Serial number of the instrument
•
QR code for device documentation
•
Manufacturer information
The type label contains the serial number of the instrument. With it
you can find the following data on our homepage:
•
Product code of the instrument (HTML)
•
Delivery date (HTML)
•
Order-specific instrument features (HTML)
•
Operating instructions at the time of shipment (PDF)
•
Safety instructions and certificates
Move to "
" and enter in the search field the serial
number of your instrument.
Alternatively, you can access the data via your smartphone:
•
Download the VEGA Tools app from the "
Apple App Store
" or the
"
Google Play Store
"
•
Scan the Data Matrix code on the type label of the instrument or
•
Enter the serial number manually in the app
3.2 Principle of operation
The VEGATOR 142 is a double-channel controller for level detection
with two 4 … 20 mA sensors. Simple monitoring and control functions
can be realised via the integrated relays. Typical applications are
monitoring functions such as overfill and dry run protection. As an
alternative to the double-channel level detection, a two-point control
can be activated.
The VEGATOR 142 controller powers the connected sensors and
simultaneously processes their measuring signals. Each input is
continuously monitored for line break or short-circuit. In addition, fault
messages delivered by the sensor are processed.
The current of a connected 4 … 20 mA sensor is measured and
evaluated. The switching point of the relay can be adjusted to any in-
Scope of delivery
Type label
Serial number
Application area
Functional principle