14
5 Connecting to power supply
VEGASWING 53 • Transistor (PNP)
61861-EN-230208
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
Warning:
Connect only in the complete absence of line voltage.
•
The electrical connection must only be carried out by trained,
qualified personnel authorised by the plant operator.
•
Always switch off power supply, before connecting or disconnect
-
ing the instrument.
The instrument is connected with standard three-wire cable without
shielding. If electromagnetic interference is expected which is above
the test values of EN 61326 for industrial areas, shielded cable should
be used.
Use cable with round cross section. Depending on the plug connec-
tion, you have to select the outer diameter of the cable respectively so
that the seal effect of the cable gland is ensured.
•
Valve plug ISO 4400, ø 4.5 … 7 mm
•
Valve plug ISO 4400 with IDC crimping technology, ø 5.5 … 8 mm
Use cable with a round wire cross section and tighten the cable gland.
Make sure that the cable used has the required temperature resist-
ance and fire safety for max. occurring ambient temperature
When mounting outdoors, on cooled vessels or in moist areas in
which cleaning is made with steam or high pressure, the sealing of
the cable gland is very important.
Note safety instructions
Connection cable
Cable glands