VEGASOURCE 86 • Source holder
1008685-EN-230810
Version with pneumatic switching ON/OFF
The pneumatic actuator must not be used under ambient conditions
that can cause corrosion in and on the pneumatic actuator.
4.2 Mounting instructions
For continuous level measurement the source holder must be
mounted slightly above or at the height of the max. level. The radiation
must be directed exactly towards the detector mounted on the op-
posite side.
The source holder VEGASOURCE should be mounted as close as
possible to the vessel.
However, with large measuring ranges and small vessel diameters, a
gap can often not be avoided.
If there are gaps or empty spaces around the installation, provide
protective fences or grids to keep hands away from the dangerous
area. Such areas must be marked accordingly.
45
°
30
°
Fig. 8: Orientation - Source holder
Beam angle 30° and 45°
The source holder must be included in the potential equalization of
the plant. For this purpose, the mounting bracket must be electrically
connected to ground potential.
To ensure a good electrical contact between the source holder and
the mounting bracket, the lock washers must be used according to
the following illustration.
Use the prescribed torque for the mounting screws. Make sure that
the screws have electrical contact to the potential equalization.
Orientation - Level meas-
urement
Screw locking device
Summary of Contents for VEGASOURCE 86
Page 1: ...Operating Instructions VEGASOURCE 86 Source holder Document ID 1008685...
Page 3: ...3 Contents VEGASOURCE 86 Source holder 1008685 EN 230810 Editing status 2023 08 10...
Page 49: ...49 10 Supplement VEGASOURCE 86 Source holder 1008685 EN 230810 10 3 Manufacturer declaration...
Page 52: ...52 Notes VEGASOURCE 86 Source holder 1008685 EN 230810...
Page 53: ...53 Notes VEGASOURCE 86 Source holder 1008685 EN 230810...
Page 54: ...54 Notes VEGASOURCE 86 Source holder 1008685 EN 230810...
Page 55: ...55 Notes VEGASOURCE 86 Source holder 1008685 EN 230810...