26
6 Setup
VEGASOURCE 86 • Source holder
1008685-EN-230810
6 Setup
6.1 Adjustment - Version X, C - Manual switching
ON/OFF
Warning:
Before switching on the radiation, make sure that no persons are
inside the radiated areas (also not inside the vessel).
Radiation must only be switched on by trained personnel.
The figures in brackets refer to the following illustration.
1
2
3
4
Fig. 14: Switching the radiation on - VEGASOURCE 86 version X, C
1 Loosen safety screw
2 Fold up the safety screw
3 Turn stirrup by 180°
4 Fold up and tighten safety screw.
1. Initial situation: Source holder is in position OFF
2. Open and remove padlock
Note:
The combination lock can be opened with the number combination
000
on delivery.
We recommend changing the number combination to a personal
number sequence. Make a note of the number combination and keep
it in a safe place.
3. Loosen safety screw (1)
4. Fold up the safety screw (2)
5. Turn the stirrup clockwise by 180° (3)
Switching the radiation