83
10 Supplement
VEGAPULS 69 • Modbus and Levelmaster protocol
47252-EN-210205
Parameter
Length
Code/Data
Request:
Function Code
1 Byte
0x10
Start Address
2 Bytes
0x0000 to 0xFFFF
Register Value
2 Bytes
0x0001 to 0x007B
Byte Number
1 Byte
2*N
Register Value
N*2 Bytes
Data
Response:
Function Code
1 Byte
0x10
Sub Function Code
2 Bytes
0x0000 to 0xFFFF
Data
2 Bytes
0x01 to 0x7B
FC17 Report Sensor ID
With this function code, the sensor ID on Modbus is queried.
Parameter
Length
Code/Data
Request:
Function Code
1 Byte
0x11
Response:
Function Code
1 Byte
0x11
Byte Number
1 Byte
Sensor ID
1 Byte
Run Indicator Status
1 Byte
FC43 Sub 14, Read Device Identification
With this function code, the Device Identification is queried.
Parameter
Length
Code/Data
Request:
Function Code
1 Byte
0x2B
MEI Type
1 Byte
0x0E
Read Device ID Code
1 Byte
0x01 to 0x04
Object ID
1 Byte
0x00 to 0xFF
Response:
Function Code
1 Byte
0x2B
MEI Type
1 Byte
0x0E
Read Device ID Code
1 Byte
0x01 to 0x04
Confirmity Level
1 Byte
0x01, 0x02, 0x03, 0x81, 0x82, 0x83
More follows
1 Byte
00/FF
Next Object ID
1 Byte
Object ID number
Number of Objects
1 Byte
List of Object ID
1 Byte
List of Object length
1 Byte
List of Object value
1 Byte
Depending on the Object ID