61
8 Diagnosis, asset management and service
VEGAPULS 69 • Modbus and Levelmaster protocol
47252-EN-210205
Out of specification:
The measured value is unreliable because an
instrument specification was exceeded (e.g. electronics temperature).
This status message is inactive by default.
Maintenance required:
Due to external influences, the instrument
function is limited. The measurement is affected, but the measured
value is still valid. Plan in maintenance for the instrument because a
failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default.
Code
Text message
Cause
Rectification
F013
no measured value
available
Sensor does not detect an echo during
operation
Antenna system dirty or defective
Check or correct installation and/or pa
-
rameter settings
Clean or exchange process component
or antenna
F017
Adjustment span too
small
Adjustment not within specification
Change adjustment according to the limit
values (difference between min. and max.
≥ 10 mm)
F025
Error in the lineariza
-
tion table
Index markers are not continuously rising,
for example illogical value pairs
Check linearization table
Delete table/Create new
F036
No operable software
Failed or interrupted software update
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
F040
Error in the electronics
Hardware defect
Exchanging the electronics
Send instrument for repair
F080
General software error
General software error
Disconnect operating voltage briefly
F105
Determine measured
value
The instrument is still in the switch-on
phase, the measured value could not yet
be determined
Wait for the end of the switch-on phase
Duration up to approx. 3 minutes depend-
ing on the version and parameter settings
F113
Communication error
EMC interference
Remove EMC influences
F125
Impermissible electron-
ics temperature
Temperature of the electronics in the non-
specified range
Check ambient temperature
Insulate electronics
Use instrument with higher temperature
range
F260
Error in the calibration
Error in the calibration carried out in the
factory
Error in the EEPROM
Exchanging the electronics
Send instrument for repair
F261
Error in the instrument
settings
Error during setup
False signal suppression faulty
Error when carrying out a reset
Repeat setup
Carry out a reset
Failure