6 Set up the sensor with the display and adjustment module
VEGAPULS 69 • Modbus and Levelmaster protocol
47252-EN-210205
well as the measured value memory, is deleted. Order-related settings
are not taken over into the current parameters after this reset.
The following table shows the scope of the reset function and the
default values of the instrument:
Menu
Menu item
Default value
Setup
Measurement
loop name
Sensor
Units
Distance in m
Temperature in °C
Application
Medium: Crushed stones/gravel
Application: Silo
Vessel bottom: Flat
Vessel height/Measuring range: 120 m
Min. adjustment
120 m
Max. adjustment
0,000 m(d)
Damping
0.0 s
Current output
mode
Output characteristics: 4 … 20 mA
Fault mode: < 3.6 mA
►
Current output -
Min./Max.
Min. current: 3.8 mA
Max. current: 20.5 mA
Lock/Unlock ad
-
justment
Released
PIN: 0000
Display
Displayed value 1 Filling height
Displayed value 2 Temperature
Backlight
Switched off
Additional adjust-
ments
Date/Time
Time format: 24 h
Scaling size
Volume
l
Scaling format
100.00 lin %, 100 l
0.00 lin %, 0 l
Current output 1
and 2 size
Lin %
Current output 1
and 2 adjustment
100.00 %, 100 l
0.00 %, 0 l
Linearisation
Linear
HART mode
HART address: 0
Loop current mode: Analogue cur-
rent output
The instrument settings are copied with this function. The following
functions are available:
•
Read from sensor:
Read data from sensor and store into the
display and adjustment module
Additional settings - Copy
instrument settings