6 Setup
VEGASOURCE 82 • Source holder
66076-EN-220119
1
2
Fig. 27: Position indication with protective cap - VEGASOURCE 82 Version X, C
1 Radiation OFF
2 Radiation ON
Version X
With version X, the protective cap can be secured with a padlock in
OFF position
Version C
With version C, the protective cap can be secured with a padlock in
OFF or ON position
6.2 Adjustment - Version B - Pneumatic
switchover
Warning:
Before switching on the radiation, make sure that no persons are
inside the radiated areas (also not inside the vessel).
Radiation must only be switched on by trained personnel.
Note:
The pneumatic actuator must not be used under ambient conditions
that can cause corrosion in and on the pneumatic actuator.
Initial situation: Pneumatic switchover is connected correctly. You can
find the required pressure values under "
Technical data
" in the annex.
The source holder is in AUS-OFF position.
Switching the radiation
on