Setup and Operation Manual
Calibration Error Codes
58
E60 - STOP ADVANCE ERROR
The volume for flow control relay 2 is set to a volume greater than the volume entered for flow control relay 1. Fix
this condition by setting the volume for SET ADVANCE STOP to a number of units smaller than the volume for
SET SLOW FLOW.
Calibration Error Codes
Calibration errors occur during meter calibration and must be corrected in order to continue with this process.
E64 - FUEL TYPE UNDEFINED
A meter calibration must be performed for this product.
E65 - CALIBRATION ERROR
One of the multipoint calibration values is different by more that 0.25% of the previous value. For example if
calibration point 1 is 100.00 than calibration point 2 must be between 99.75 and 100.25 counts. Push any button
to clear the error code. Push the NEXT button to begin calibrating a new flow rate point using either Auto
Calibrate or Multicalibrate. It is also acceptable to calculate the counts and enter them manually using the manual
calibrate feature.
Other Problems
SYSTEM BOOTS UP OKAY, BUT DISPLAY KEYS DON’T RESPOND
Electrical interference or noise could cause this problem. Make sure the IB and the Display Head each have a
good ‘Earth Ground’ and the ground/shield between the IB and Display Head is connected. Try to find the source
by turning on and off motors/pumps and other equipment in the area. Make sure this equipment is well grounded.