Figure 4-8. Temperature Control Mode Selection Screen
Note:
When switching between PyroControl or T/C modes, the best process results are
achieved when recipes are written that allow the temperature to stabilize at the process
setpoint for 8-10 minutes prior to the beginning of deposition.
TEMPERATURE SETTINGS
Caution:
When switching between temperature modes, it is important to have a
smooth transition. This will prevent large spikes in power supply output when the switch
occurs.
To avoid temperature fluctuations when changing modes, determine the correct T/C
temperature setting that corresponds to the desired PyroControl temperature setting. Note
that these settings may not be the same and may require several calibration growth runs to
determine.