Calibration
Fine-tuning for X-Y Calibration
330
Dimension 3100 Manual
Rev. D
4. Return to the
Scanner Calibration
dialog box.
5. Select the
X fast derate
or
Y slow derate
parameter.
6. Enter the new
X fast derate
or
Y slow derate
value calculated in
scanner’s fast axis to more closely match calculated distances with actual feature distances.
7. To set the new parameter value, click
OK
.
17.7.5 Measure Vertically at 150V Scan Size
1. Select two widely-spaced features on the sample image of known separation.
2. Use the mouse to draw a vertical line between them.
Note:
For example, on a 10µm, silicon reference, draw the line from the top edge of
one pit to the top edge of another pit as far away as possible. The microscope
displays the measured distance next to the line.
3. Verify that the microscope’s measured distance agrees with the known vertical distance. If
there is significant disagreement between the two, execute the fine tuning procedure; go to
the next step. If the displayed distance agrees with the known distance, no further calibration
is required.
4. Adjust the
Yslow derate
value by using one of two methods: 1) the
or 2) the
Measure Horizontally at 150 V Scan Size:
17.7.6 Change Scan angle and Repeat Calibration Routines
•
Change the
Scan angle
on the
Scan Controls
panel to
90
º.
•
Repeat steps above for the following parameters:
Y fast sens
,
X slow sens
,
Y fast der
,
and
X slow der
to ensure the scanner is calibrated properly along the X- and Y-axis for
scanning at 90º.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......