Force Imaging
Force Calibration Control Panels and Menus
Rev. D
Dimension 3100 Manual
215
13.3.3 Channel 1, 2, 3 Panels
Data Type
This parameter allows you to select the type of data you want to display on the vertical axis.
Data Scale
This parameter sets the vertical scale in the force plot. Increasing this parameter expands the range
of the display about the centerline causing more of the force curve to fall on the graph. Initially, set
this parameter to its maximum value, then gradually reduce.
Data Center
This parameter shifts the displayed force plot in the vertical axis.
Note:
For
Data type
=
Deflection
in Contact Mode or
Data type
=
Amplitude
in
TappingMode, the data is also shifted by the setpoint.(i.e., if the
Data type
=
Deflection
and
Data Center
=
0
, then the setpoint is at the center of the vertical
scale.
Deflection Sensitivity
This item relates the cantilever deflection signal in volts from the controller to the actual deflection
of the cantilever (in nm). It equals the inverse of the slope of the deflection versus Z piezo position
line when the tip is in contact with a very hard, stiff sample as shown in
NanoScope system automatically calculates and enters the value from the graph once you use the
mouse to draw a line parallel to the graph. You must properly determine the deflection sensitivity
value before deflection data in nanometers can accurately display. For a proper force curve, the line
has a negative slope with typical values of 10-100 nm/V; however, by convention, values appear as
positive in the menu.
Amplitude Sensitivity (TappingMode)
Amplitude Sensitivity
relates the vibrational amplitude of the cantilever (in nm) to the amplitude
signal (in V) from the controller. To calculate the
Amplitude Sensitivity
, measure the slope of the
RMS amplitude
versus the
Z piezo position
when the tip is in contact with a very hard, stiff
sample. The NanoScope software automatically calculates and enters the value from the graph after
you use the mouse to draw a line to the graph. You must properly calibrate the
Amplitude
Sensitivity
before amplitude data can be correctly displayed in nm.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......