Single Scan Operation
4-9
Figure 4-8: All Units Dialog Box
3
Make your changes to the
Height (Z)
and
Notation
settings.
4
Click
Apply
, and then click
OK
.
REFERENCE/MEASUREMENT CURSORS
The reference (
R
) cursor and measurement (
M
) cursor define the portion of the profile trace for
leveling or performing analytical functions. You can adjust the bandwidth at each cursor to average
the data points within the cursor’s bandwidth. This is useful for leveling and average step height
measurements.
NOTE –
Veeco recommends that you do
not
use zero (0) bandwidths unless you want point-
to-point measurements. Such measurements can vary substantially from the averaged results
that you get with non-zero bandwidths.
BASIC CURSOR POSITIONING
Cursor positioning is critical for obtaining accurate results. The simplest way to reposition the cursors
is to use the mouse to drag the
R
and
M
cursor flags at the top of the cursors to new positions in the
data plot.
Selecting both
R
and
M
cursors allows you to reposition the cursors while maintaining the same
distance between them. Hold down the
C
TRL
key while dragging with the left mouse button on the
R
or
M
cursor flag. Now dragging either flag causes both cursors to move simultaneously.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......